top of page
Troubleshooting AMX
Error Number
Error Code / System Message
Error Question
What does low or zero 24 V on the Definium AMX 700 debug header indicate?
Short Answer
The detector may be unregistered, incorrectly configured, or unable to communicate through the wireless access point.
Possible Causes
Detector not registered; AP configuration issue; unsupported wireless country/channel combination.
Troubleshooting Steps
Check the registered-detector list. Register the detector if needed, then insert it in the proper bin to pair. If AP configuration messages appear, verify the selected country configuration and AP operation.
Parts Areas to Check
bottom of page