Troubleshooting AMX

Error Number
BATTERY TOO HIGH
Error Code / System Message
​
Error Question
How do I troubleshoot this AMX-4 Series issue: “BATTERY TOO HIGH” Message?
Short Answer
This Usually Indicates That The Unit Was Attempted To Be Operated (Rotor And Exposure) Too Soon After A Charging Cycle. In Normal Applications, When The Unit Is Unplugged, It is Normally Driven To Another Location (More Than 50 Feet), Thus Giving The Batteries Time To “Level Off”.
Possible Causes
Problem: “BATTERY TOO HIGH” Message
Troubleshooting Steps
This Usually Indicates That The Unit Was Attempted To Be Operated (Rotor And Exposure) Too Soon After A Charging Cycle. In Normal Applications, When The Unit Is Unplugged, It is Normally Driven To Another Location (More Than 50 Feet), Thus Giving The Batteries Time To “Level Off”.
Parts Areas to Check
Battery; Rotor