Troubleshooting AMX
Error Number
BATTERY TOO LOW - RECHARGE REQUIRED
Error Code / System Message
BATTERY TOO LOW - RECHARGE REQUIRED
Error Question
What does “BATTERY TOO LOW - RECHARGE REQUIRED” mean on an AMX-Series unit?
Short Answer
It indicates the battery capacity/voltage is too low for reliable operation and the unit should be recharged before continued clinical use.
Possible Causes
Low state of charge; aged battery capacity; capacity-meter synchronization issue.
Troubleshooting Steps
Recharge the unit fully. If the message returns quickly, verify battery condition under load and check capacity-meter synchronization before replacing the battery set.
Parts Areas to Check
Battery set; capacity meter