top of page

Troubleshooting AMX

Error Number
F93 On Rotor Board Opens
Error Code / System Message
​
Error Question
How do I troubleshoot this AMX-4 Series issue: F93 On Rotor Board Opens When A New Filament kVp Board (46-264986G2) Is Installed, FETs On Rotor Board NOT Blown?
Short Answer
J8 Connector On Filament kVp Board Is Mis-keyed. Re-orient Key On J8 (FIL/kVp Board Side). CQA 972175
Possible Causes
11) Problem: F93 On Rotor Board Opens When A New Filament kVp Board (46-264986G2) Is Installed, FETs On Rotor Board NOT Blown
Troubleshooting Steps
J8 Connector On Filament kVp Board Is Mis-keyed. Re-orient Key On J8 (FIL/kVp Board Side). CQA 972175
Parts Areas to Check
Filament kVp Board; Rotor
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page