Troubleshooting AMX
Error Number
kVp Output Waveform Initial Rise 10-20 kVp High For 4-5 ms. (During Calibration And Applications), And Then Drops To Proper Selected kVp Value. NO System Error Codes Displayed During Applications Other Than Occasional “k
Error Code / System Message
Error Question
How do I troubleshoot this AMX-4 Series issue: kVp Output Waveform Initial Rise 10-20 kVp High For 4-5 ms. (During Calibration And Applications), And Then Drops To Proper Selected kVp Value. NO System Error Codes Displayed During Applications Other Than Occasional “k?
Short Answer
Replace bad Power Module SCR Q3 Block and/or Filament kVp Board
Possible Causes
Problem: kVp Output Waveform Initial Rise 10-20 kVp High For 4-5 ms. (During Calibration And Applications), And Then Drops To Proper Selected kVp Value. NO System Error Codes Displayed During Applications Other Than Occasional “kVp LIMITS” Error During Calibration.
Troubleshooting Steps
Replace bad Power Module SCR Q3 Block and/or Filament kVp Board
Parts Areas to Check
Filament kVp Board; Power Module SCR Q3
For Related Models