Troubleshooting AMX
Error Number
kVp’s High In Applications And During And During kVp Calibration. In The “Check- Sum Bypass Mode” (With Defaults Loaded), Tap Relays (80 kVp Selected) K2 And K4 not selected.
Error Code / System Message
Error Question
How do I troubleshoot this AMX-4 Series issue: kVp’s High In Applications And During And During kVp Calibration. In The “Check- Sum Bypass Mode” (With Defaults Loaded), Tap Relays (80 kVp Selected) K2 And K4 not selected?
Short Answer
Replace the defective Battery Charger Board (J1-5: “Battery V And Charger Currents” Output Improper). NOTE: System Parameters Are Still “Looked At” In The “Check Sum By-pass” Mode.
Possible Causes
Problem: kVp’s High In Applications And During And During kVp Calibration. In The “Check- Sum Bypass Mode” (With Defaults Loaded), Tap Relays (80 kVp Selected) K2 And K4 not selected.
Troubleshooting Steps
Replace the defective Battery Charger Board (J1-5: “Battery V And Charger Currents” Output Improper). NOTE: System Parameters Are Still “Looked At” In The “Check Sum By-pass” Mode.
Parts Areas to Check
Battery Charger Board; Battery; Tap Relay
For Related Models
AMX-4, AMX-4+, Definium AMX 700