top of page

Troubleshooting AMX

Error Number
No “normal” kVp waveform can be viewed (Applications Or Check-Sum Bypass) With Errors 439 Or 469
Error Code / System Message
​
Error Question
How do I troubleshoot this AMX-4 Series issue: No “normal” kVp waveform can be viewed (Applications Or Check-Sum Bypass) With Errors 439 Or 469?
Short Answer
Temporarily Jumper TP-14 To TP-15 On Filament kVp Board To Obtain A “normal” kVp Waveform
Possible Causes
16) Problem: No “normal” kVp waveform can be viewed (Applications Or Check-Sum Bypass) With Errors 439 Or 469
Troubleshooting Steps
Temporarily Jumper TP-14 To TP-15 On Filament kVp Board To Obtain A “normal” kVp Waveform
Parts Areas to Check
Filament kVp Board
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page