Troubleshooting AMX
Error Number
Open filament in X-Ray Tube (Error 45F) during PREP. Defective Filament kVp Control Board (Q170, Q160 or Q173 shorted, Inverter running at maximum output. CQA 964232.
Error Code / System Message
45F
Error Question
How do I troubleshoot this AMX-4 Series issue: Open filament in X-Ray Tube (Error 45F) during PREP. Defective Filament kVp Control Board (Q170, Q160 or Q173 shorted, Inverter running at maximum output. CQA 964232?
Short Answer
Replace Filament kVp Board
Possible Causes
3) Problem: Open filament in X-Ray Tube (Error 45F) during PREP. Defective Filament kVp Control Board (Q170, Q160 or Q173 shorted, Inverter running at maximum output. CQA 964232.
Troubleshooting Steps
Replace Filament kVp Board
Parts Areas to Check
Filament kVp Board; X-Ray Tube
For Related Models
AMX-4, AMX-4+, Definium AMX 700