top of page
Troubleshooting AMX
Error Number
kV/mA cannot be adjusted from AMX or UIF
Error Code / System Message
Error Question
Why can’t I adjust kV or mA on the Definium AMX 700?
Short Answer
The functional test expects kV and mA changes from both the AMX controls and the user interface to track correctly.
Possible Causes
PC protocol/control communication issue; Beetle Board or AMX control-path issue.
Troubleshooting Steps
Allow the PC to boot, enter an exam and protocol, then adjust kV/mA from the AMX and UIF. If only one control path fails, isolate that interface; if both fail, check PC/Beetle communication.
Parts Areas to Check
UIF; Beetle Board; AMX controller; PC
For Related Models
Definium AMX 700