top of page

Troubleshooting AMX

Error Number
MPPS updates not sent
Error Code / System Message
​
Error Question
Why can’t the Definium AMX 700 send MPPS updates?
Short Answer
Verify networking and the MPPS server IP address, case-sensitive AE Title and DICOM port, then run C-Echo.
Possible Causes
Network issue; MPPS configuration mismatch; server unavailable.
Troubleshooting Steps
Open Network Connections, select the MPPS device under HIS/RIS, verify IP/AE Title/port, run C-Echo and review the system error log if the test fails.
Parts Areas to Check
MPPS server; AMX 700 DICOM configuration
For Related Models
Definium AMX 700
bottom of page