top of page

Troubleshooting AMX

Error Number
Shutting Down
Error Code / System Message
Shutting Down... - Expected display during orderly key-switch shutdown
Error Question
What does “Shutting Down...” mean on a GE Definium AMX 700?
Short Answer
Shutting Down...: Expected display during orderly key-switch shutdown.
Possible Causes
Subsystem/category: PC / Beetle Communication. Source classification: STATUS.
Troubleshooting Steps
If this does not appear, check key-switch scaled signal and IF_PS_CTRL behavior.
Parts Areas to Check
Beetle Board; Digital PSA; key switch
For Related Models
Definium AMX 700
bottom of page