Troubleshooting AMX
Error Number
Error 326
Error Code / System Message
326
Error Question
What should I check for error 326 on the AMX-4 Series?
Short Answer
1) Check items 1, 2, and 3 from Error 322 2) Batteries may be defective If Voltage Is Really More Than 150 volts Note: If unit does not charge (no voltage across sense resistor AMX1 A3 R1) - verify that the safety contactor contacts are not welded shut
Possible Causes
1) Check items 1, 2, and 3 from Error 322 2) Batteries may be defective If Voltage Is Really More Than 150 volts Note: If unit does not charge (no voltage across sense resistor AMX1 A3 R1) - verify that the safety contactor contacts are not welded shut
Troubleshooting Steps
1) Check items 1, 2, and 3 from Error 322 2) Batteries may be defective If Voltage Is Really More Than 150 volts Note: If unit does not charge (no voltage across sense resistor AMX1 A3 R1) - verify that the safety contactor contacts are not welded shut
Parts Areas to Check
Safety Contactor
For Related Models
AMX-4, AMX-4+, Definium AMX 700