Troubleshooting AMX

Error Number
Error 467
Error Code / System Message
467
Error Question
What should I check for error 467 on the AMX-4 Series?
Short Answer
1. Capacitor C5 Leaky AMX-4 Units (Does Not Apply To AMX-4+ Units) 2. Defective CPU Board 3. Bad battery Cell(s) 4. Defective Filament kVp Control Board 5. Unit recalibration is required (If In The “CHECKSUM-BYPASS” Mode) 6. HV Cable Open/Shorted Internally Or Poor Pin Spread On HV Candles
Possible Causes
1. Capacitor C5 Leaky AMX-4 Units (Does Not Apply To AMX-4+ Units) 2. Defective CPU Board 3. Bad battery Cell(s) 4. Defective Filament kVp Control Board 5. Unit recalibration is required (If In The “CHECKSUM-BYPASS” Mode) 6. HV Cable Open/Shorted Internally Or Poor Pin Spread On HV Candles
Troubleshooting Steps
1. Capacitor C5 Leaky AMX-4 Units (Does Not Apply To AMX-4+ Units) 2. Defective CPU Board 3. Bad battery Cell(s) 4. Defective Filament kVp Control Board 5. Unit recalibration is required (If In The “CHECKSUM-BYPASS” Mode) 6. HV Cable Open/Shorted Internally Or Poor Pin Spread On HV Candles
Parts Areas to Check
CPU Board; Battery
For Related Models
AMX-4, AMX-4+, Definium AMX 700