top of page

Troubleshooting AMX

Error Number
Fuse 2A blown
Error Code / System Message
2A fuse
Error Question
How do I identify a blown 2A fuse on the Definium AMX 700 Digital Power Supply?
Short Answer
With proper input voltage present, no voltage at debug-header connection 6 indicates the 2A fuse is the likely cause.
Possible Causes
Blown 2A fuse.
Troubleshooting Steps
Verify proper input voltage is present, then measure debug-header connection 6. If the reading is absent, inspect/replace the 2A fuse using the correct fuse orientation.
Parts Areas to Check
Digital Power Supply; 2A fuse; 12 V converter input branch
For Related Models
Definium AMX 700
bottom of page