Troubleshooting AMX

Error Number
Capacity bar graph disagrees with battery voltage
Error Code / System Message
CHARGE COMPLETE
Error Question
How do I synchronize the AMX-Series battery capacity meter with the displayed capacity?
Short Answer
The manual provides two methods: complete a full charge to CHARGE COMPLETE, or use the service LOOP TEST procedure to copy internal capacity data to the displayed capacity.
Possible Causes
Capacity-display drift after charging or battery replacement; stale internal capacity data.
Troubleshooting Steps
Prefer a full charge through CHARGE COMPLETE. If service synchronization is required, use the manual's service procedure to synchronize internal and displayed capacity.
Parts Areas to Check
Capacity meter; CPU non-volatile memory; batteries
For Related Models