top of page

Troubleshooting AMX

Error Number
Hand switch not working
Error Code / System Message
​
Error Question
How do I test the Brivo XR285amx wired hand switch?
Short Answer
The Wired Hand Switch diagnostic tests the X-ray generation initiation circuitry.
Possible Causes
Hand-switch or initiation-circuit fault.
Troubleshooting Steps
Run Wired Hand Switch in SUIF and follow the on-screen test. If it fails, isolate the hand switch, cable, and system-control initiation path.
Parts Areas to Check
​
For Related Models
Brivo XR285amx