top of page

Troubleshooting AMX

Error Number
Djinn inverter issues
Error Code / System Message
​
Error Question
How do I test the Djinn inverter on the Brivo XR285amx?
Short Answer
The Inverter Test drives the MOSFET gates with a low DC bus to verify inverter functionality.
Possible Causes
Inverter/power-module fault; resonant component issue; DC-bus problem.
Troubleshooting Steps
Run Inverter Test and Resonant Components. If either fails, review kV/BB errors and isolate the Power Module, resonant components, and DC-bus connections.
Parts Areas to Check
​
For Related Models
Brivo XR285amx
bottom of page