Troubleshooting AMX
Error Number
kVp calibration problem
Error Code / System Message
KVP CAL ERROR 1/2/4; X-RAY BYTE/WORD LIMIT
Error Question
How do I troubleshoot AMX-Series kVp calibration errors?
Short Answer
kVp calibration errors occur when required kVp/mA combinations cannot be reached, calibration limits are exceeded, or X-ray feedback/calibration data is inconsistent.
Possible Causes
Incorrect battery voltage; HV divider/feedback issue; filament/tap calibration issue; generator output limit.
Troubleshooting Steps
Verify battery voltage, calibration meter setup, kVp feedback/divider wiring, filament/tap calibration, and the applicable kVp points. Use the specific KVP CAL or X-RAY LIMIT message.
Parts Areas to Check
HV divider; filament/kVp board; generator; battery
For Related Models
AMX-4, AMX-4+, Definium AMX 700