top of page
Troubleshooting AMX
Error Number
Unit will not drive normally
Error Code / System Message
ERROR 12/13/19/210-218
Error Question
How do I troubleshoot an AMX-Series drive-control problem?
Short Answer
Use the run-time drive display to compare left/right drive command and feedback while observing handle displacement and safety interlocks.
Possible Causes
Handle calibration or Hall-sensor issue; feedback mismatch; drive-control board; relay/fuse/interlock fault.
Troubleshooting Steps
Check handle calibration, command versus feedback values, tube-park/bumper/drive-enable inputs, drive fuses and motor relays. Use the specific application error code when present.
Parts Areas to Check
Drive handle; Hall sensors; drive control board; motor relays; drive fuses
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page