Short Answer
The manual recommends performing a new filament calibration first and then continuing with filament-driver diagnostics if the error persists.
Possible Causes
Filament calibration/aging; filament driver issue; measurement path issue; tube issue.
Troubleshooting Steps
Run Filament Drive Calibration. During calibration, some low/high mA messages can occur and may be acceptable if calibration completes successfully. If the problem persists afterward, follow the filament driver and mA measurement diagnostic sequence.