top of page

Troubleshooting AMX

Error Number
FlashPad detector not register
Error Code / System Message
​
Error Question
How is a FlashPad detector registered on XR220amx or digitally upgraded XR200amx?
Short Answer
Initial detector registration is performed with a tether connection so the host PC can recognize the detector in relation to the system.
Possible Causes
Missing tether; detector not powered/booted; detector or system communication problem.
Troubleshooting Steps
Connect the tether, power the detector, open Detector Registration and use Add/Edit as required. If registration fails, run BIST/Ping and compare alternate communication methods.
Parts Areas to Check
​
For Related Models
Optima XR200amx, Optima XR220amx