Oscilloscope triggering during calibration
Error Code / System Message
Error Question
How should I trigger an oscilloscope for AMX-Series generator calibration?
Short Answer
The manual recommends establishing the scope trigger before entering CALIBRATE KVP because the unit fails power-up tests in calibration mode.
Possible Causes
Difficult oscilloscope triggering during service calibration.
Troubleshooting Steps
Set up the trigger in normal operating mode using the specified low-kVp/mAs test exposure, then enter calibration with the scope already triggering reliably.
Parts Areas to Check
Generator calibration test points
For Related Models
AMX-4, AMX-4+, Definium AMX 700