top of page
Troubleshooting AMX
Error Number
Battery meter showed 100%
Error Code / System Message
Error Question
What should I check when the XR240amx battery meter shows 100%?
Short Answer
A 100% meter alone is not the full battery-health assessment; the manual also references battery current, LVLE2 current, pack voltage, and temperature.
Possible Causes
Surface charge; weak battery pack; abnormal current or pack voltage.
Troubleshooting Steps
Use Power Control Limits diagnostics to review individual pack voltages, battery current, LVLE2 current and temperatures. Remove surface charge as instructed before comparing readings.
Parts Areas to Check
For Related Models
Optima XR240amx
bottom of page