Troubleshooting AMX

Error Number
X-ray terminates early
Error Code / System Message
ERROR 462 / 466 / 467 / 468
Error Question
What can cause an AMX-Series exposure to terminate early?
Short Answer
The manual lists excessive exposure time, hardware backup timer expiration, explicit exposure-stop state, interlock changes, or generator feedback faults as possible causes.
Possible Causes
Timing limit reached; stop command; interlock opened; generator output/feedback mismatch.
Troubleshooting Steps
Review ERROR 462, 466, 467, 468 and related generator codes. Check exposure time, interlocks, timing backup circuits, command/feedback status, and battery voltage under load.
Parts Areas to Check
Generator timing; CPU; interlocks; battery condition
For Related Models
AMX-4, AMX-4+, Definium AMX 700