top of page
Troubleshooting AMX
Error Number
Error 0x0100
Error Code / System Message
Error Question
What does 0x0100 mean on a GE Brivo XR285amx?
Short Answer
0x0100: Unknown FW error.
Possible Causes
The Djinn Beetle generator reported a Heater 1 FW fault: Unknown FW error.
Troubleshooting Steps
Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx