top of page

Troubleshooting AMX

Error Number
Error 0x0103
Error Code / System Message
​
Error Question
What does 0x0103 mean on a GE Brivo XR285amx?
Short Answer
0x0103: Demand too high (target > max heater).
Possible Causes
The Djinn Beetle generator reported a Heater 1 FW fault: Demand too high (target > max heater).
Troubleshooting Steps
Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.
Parts Areas to Check
​
For Related Models
Brivo XR285amx
bottom of page