top of page

Troubleshooting AMX

Error Number
Error 0x0200
Error Code / System Message
​
Error Question
What does 0x0200 mean on a GE Brivo XR285amx?
Short Answer
0x0200: Unkwn FW error.
Possible Causes
The Djinn Beetle generator reported a Heater 2 FW fault: Unkwn FW error.
Troubleshooting Steps
Run Fil 2 Drv Clsd Loop, Fil 2 Xfmr Primary, and Fil 2 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.
Parts Areas to Check
​
For Related Models
Brivo XR285amx