top of page
Troubleshooting AMX
Error Number
Error 0x0585
Error Code / System Message
Error Question
What does 0x0585 mean on a GE Brivo XR285amx?
Short Answer
0x0585: Input current detected while BB is OFF.
Possible Causes
The Djinn Beetle generator reported a BB SW fault: Input current detected while BB is OFF.
Troubleshooting Steps
Run Rise Time/Regulation, Power Supplies, and relevant DC-bus checks. Verify battery input is at least 144 VDC, inspect the DC-bus cable and breaker path, and confirm the 48 V supply is within specification.
Parts Areas to Check
For Related Models
Brivo XR285amx
bottom of page