top of page

Troubleshooting AMX

Error Number
Error 0x0601
Error Code / System Message
​
Error Question
What does 0x0601 mean on a GE Brivo XR285amx?
Short Answer
0x0601: Exposure logic error.
Possible Causes
The Djinn Beetle generator reported a kV FW fault: Exposure logic error.
Troubleshooting Steps
Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.
Parts Areas to Check
​
For Related Models
Brivo XR285amx
bottom of page