What does 0x0606 mean on a GE Brivo XR285amx?
Short Answer
0x0606: Too many spits.
Possible Causes
The Djinn Beetle generator reported a kV FW fault: Too many spits.
Troubleshooting Steps
Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated.