top of page

Troubleshooting AMX

Error Number
Error 0x0787
Error Code / System Message
​
Error Question
What does 0x0787 mean on a GE Brivo XR285amx?
Short Answer
0x0787: Single Wire communication impossible.
Possible Causes
The Djinn Beetle generator reported a Thermal SW fault: Single Wire communication impossible.
Troubleshooting Steps
Run Power Supply Thermal and the applicable thermal sensor tests. Check cooling, fans, sensor wiring, tube thermal-switch resistance, and single-wire communication before replacing a module.
Parts Areas to Check
​
For Related Models
Brivo XR285amx
bottom of page