top of page
Troubleshooting AMX
Error Number
Error 0x0908
Error Code / System Message
Error Question
What does 0x0908 mean on a GE Brivo XR285amx?
Short Answer
0x0908: FPGA not compatible with application software.
Possible Causes
The Djinn Beetle generator reported a Boot SW fault: FPGA not compatible with application software.
Troubleshooting Steps
Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.
Parts Areas to Check
For Related Models
Brivo XR285amx
bottom of page