top of page

Troubleshooting AMX

Error Number
Error 0x090C
Error Code / System Message
​
Error Question
What does 0x090C mean on a GE Brivo XR285amx?
Short Answer
0x090C: ST10 watchdog error.
Possible Causes
The Djinn Beetle generator reported a Boot SW fault: ST10 watchdog error.
Troubleshooting Steps
Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.
Parts Areas to Check
​
For Related Models
Brivo XR285amx
bottom of page