top of page
Troubleshooting AMX
Error Number
Error 0x0A81
Error Code / System Message
Error Question
What does 0x0A81 mean on a GE Brivo XR285amx?
Short Answer
0x0A81: Internal power supplies out of specification.
Possible Causes
The Djinn Beetle generator reported a Various SW fault: Internal power supplies out of specification.
Troubleshooting Steps
Run System Sanity and Power Supplies, then review the error log. Check the indicated internal power, NVRAM/RTC, FPGA communication, or reboot condition before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
bottom of page