top of page
Troubleshooting AMX
Error Number
Error 0x0B87
Error Code / System Message
Error Question
What does 0x0B87 mean on a GE Brivo XR285amx?
Short Answer
0x0B87: mA out of range.
Possible Causes
The Djinn Beetle generator reported a mA SW fault: mA out of range.
Troubleshooting Steps
Run mA Measurement and mA Test. Reseat the HV Tank J6-to-Power Module J7 measurement cable and perform the specified continuity/resistance checks; use filament ageing clearing/calibration when 0x0B84-0x0B86 is involved.
Parts Areas to Check
For Related Models
Brivo XR285amx
bottom of page