top of page
Troubleshooting AMX
Error Number
BATTERY ERROR 1
Error Code / System Message
BATTERY ERROR 1 - Battery calibration counter overflow
Error Question
What does BATTERY ERROR 1 mean on a GE AMX-Series mobile X-ray?
Short Answer
BATTERY ERROR 1: Battery calibration counter overflow.
Possible Causes
Battery/charger calibration count exceeded its expected range.
Troubleshooting Steps
Verify charger-board frequency/input and repeat battery calibration.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
bottom of page