Troubleshooting AMX
Error Number
CAL TAP ERROR 1
Error Code / System Message
CAL TAP ERROR 1 - Proper generator mA could not be reached
Error Question
What does CAL TAP ERROR 1 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL TAP ERROR 1: Proper generator mA could not be reached.
Possible Causes
Changing the kVp/tap could not achieve the required mA.
Troubleshooting Steps
Check mA calibration, load connections and repeat tap calibration.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700