top of page
Troubleshooting AMX
Error Number
CAL TAP ERROR 3
Error Code / System Message
CAL TAP ERROR 3 - Low-kVp tap voltage too high
Error Question
What does CAL TAP ERROR 3 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL TAP ERROR 3: Low-kVp tap voltage too high.
Possible Causes
Tube-voltage port exceeded the allowed low-kVp range.
Troubleshooting Steps
Repeat kVp calibration and inspect generator/tap wiring.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700