top of page
Troubleshooting AMX
Error Number
CAL TAP ERROR 5
Error Code / System Message
CAL TAP ERROR 5 - Filament-current feedback out of range
Error Question
What does CAL TAP ERROR 5 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL TAP ERROR 5: Filament-current feedback out of range.
Possible Causes
Filament-current digital-to-analog/feedback value was outside expected range.
Troubleshooting Steps
Check mAs/filament calibration, feedback circuitry and tube connections.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700