top of page

Troubleshooting AMX

Error Number
CHARG WORD LIMIT
Error Code / System Message
CHARG WORD LIMIT - Charger word calibration limit exceeded
Error Question
What does CHARG WORD LIMIT mean on a GE AMX-Series mobile X-ray?
Short Answer
CHARG WORD LIMIT: Charger word calibration limit exceeded.
Possible Causes
Charge-current calibration parameter exceeded word range.
Troubleshooting Steps
Calibrate charger; check CPU/charger-board connections and charge-scale-select logic.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page