top of page
Troubleshooting AMX
Error Number
CHECK SUM ERR 11
Error Code / System Message
CHECK SUM ERR 11 - Filament-current table checksum failed
Error Question
What does CHECK SUM ERR 11 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 11: Filament-current table checksum failed.
Possible Causes
Filament-current table checksum failed or was not calibrated.
Troubleshooting Steps
Calibrate filament current; if other generator checksums are present, calibrate generator first.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page