top of page
Troubleshooting AMX
Error Number
CHECK SUM ERR 7
Error Code / System Message
CHECK SUM ERR 7 - X-ray calibration word checksum failed
Error Question
What does CHECK SUM ERR 7 mean on a GE AMX-Series mobile X-ray?
Short Answer
CHECK SUM ERR 7: X-ray calibration word checksum failed.
Possible Causes
X-ray calibration words failed checksum.
Troubleshooting Steps
Calibrate generator.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page