top of page
Troubleshooting AMX
Error Number
Error 0205
Error Code / System Message
Error Question
What does error 0205 mean on a GE Optima XR240amx?
Short Answer
Error 0205: Filament inverter is off and current is not zero.
Possible Causes
Subsystem/category: Fil Large. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
For Related Models
Optima XR240amx
bottom of page