top of page
Troubleshooting AMX
Error Number
Error 0467
Error Code / System Message
Error Question
What does error 0467 mean on a GE Optima XR240amx?
Short Answer
Error 0467: Current measurement issue on rotation.
Possible Causes
Subsystem/category: Rotor. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run Rotation and Tube Rotor Impedance diagnostics. Inspect rotor communication, tube stator/harness connections, and Power/Main Auxiliary circuitry.
Parts Areas to Check
For Related Models
Optima XR240amx