top of page

Troubleshooting AMX

Error Number
Error 0480
Error Code / System Message
​
Error Question
What does error 0480 mean on a GE Optima XR240amx?
Short Answer
Error 0480: Communication failure between Power and Main Auxiliary modules.
Possible Causes
Subsystem/category: Rotor. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review related generator errors. Run Rotation and Tube Rotor Impedance diagnostics. Inspect rotor communication, tube stator/harness connections, and Power/Main Auxiliary circuitry.
Parts Areas to Check
​
For Related Models
Optima XR240amx
bottom of page