top of page

Troubleshooting AMX

Error Number
Error 0487
Error Code / System Message
​
Error Question
What does error 0487 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0487: Rotor: Service mode diagnostic communication failure.
Possible Causes
Subsystem/category: Rotor. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review related generator errors. Run Rotation and Tube Rotor Impedance diagnostics; inspect tube-stator/rotor wiring and Power/Main Auxiliary communication.
Parts Areas to Check
​
For Related Models
Optima XR200amx, Optima XR220amx
bottom of page