Troubleshooting AMX

Error Number
Error 0504
Error Code / System Message
​
Error Question
What does error 0504 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0504: Input stage: DC Bus rising time too slow during DC Bus charge [DC Bus < 200V].
Possible Causes
Subsystem/category: Input Stage / DC Bus. Manual recovery/error class: ABORT.
Troubleshooting Steps
Check Power Control Limits and generator input/DC-bus diagnostics. Evaluate system battery state, input cabling and generator input-stage hardware.
Parts Areas to Check
​
For Related Models
Optima XR200amx, Optima XR220amx
bottom of page