top of page

Troubleshooting AMX

Error Number
Error 0580
Error Code / System Message
​
Error Question
What does error 0580 mean on a GE Optima XR240amx?
Short Answer
Error 0580: DC-bus timeout when boot is started.
Possible Causes
Subsystem/category: Input Stage. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator and system-battery errors. Run Power Supplies / Power Control Limits and the applicable input-voltage or DC-bus diagnostics. Inspect the generator input-power path.
Parts Areas to Check
​
For Related Models
Optima XR240amx
bottom of page