top of page

Troubleshooting AMX

Error Number
Error 0605
Error Code / System Message
​
Error Question
What does error 0605 mean on a GE Optima XR240amx?
Short Answer
Error 0605: Maximum kV reached on cathode (>125% of cathode kV target).
Possible Causes
Subsystem/category: kV. Manual recovery/error class: ABORT.
Troubleshooting Steps
Review related generator errors. Run the applicable kV, Inverter, Resonant Components and Rise Time/Regulation diagnostics. Inspect the HV/tube cable path only under qualified service procedures.
Parts Areas to Check
​
For Related Models
Optima XR240amx
bottom of page