top of page
Troubleshooting AMX
Error Number
Error 0707
Error Code / System Message
Error Question
What does error 0707 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0707: Thermal: Power board temperature is rising.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Allow cooling when appropriate, verify airflow, and run generator thermal/70C diagnostics. Check the board or sensor identified by the code.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
bottom of page